Dependence of the stabilization of the given temperature of the applicator cryosurgical unit "Cryo-Pulse" from excess pressure in cryostats

Main Article Content

A.YA. Zharkov
Ya.V. Zharkov
M.M. Kovalenko
R.I. Yanchiy

Abstract

The experiments which allowed to define the optimal values of surplus pressure in cryostat by the universal automated cryosurgical device "Cryo-Pulse" which provide minimal deviation of the applicator working surface temperature from a set value of minus 1800С, minus 1000С, 00С are described. Recommendations on determination of the optimum surplus pressure for every applicator and every temperature level of cryoinfluence are given

Article Details

How to Cite
Zharkov, A. ., Zharkov, Y. ., Kovalenko, M. ., & Yanchiy, R. . (2011). Dependence of the stabilization of the given temperature of the applicator cryosurgical unit "Cryo-Pulse" from excess pressure in cryostats. Electronics and Communications, 16(1), 133–137. https://doi.org/10.20535/2312-1807.2011.16.1.274001